Section II: Equipment Specifications
Construction
Overall Dimensions:
- LFGI-3USP: 54” W (39” W at base) x 32” D x 80” H
- LFGI-4USP: 66” W (50” W at base) x 32” D x 80” H
- LFGI-6USP: 90” W (74” W at base) x 32” D x 80” H
Work Area Dimensions:
- LFGI-3USP: 34½” W x 24½” D x 29” H
- LFGI-4USP: 46½” W x 24½” D x 29” H
- LFGI-6USP: 70½” W x 24½” D x 29” H
General Info :
- Designed to fit through standard door openings and elevators.
- All stainless steel construction both inside and out with a pharmaceutical grade finish.
- 316L stainless steel work trays slide over one another and are removable to facilitate easy clean up under the work deck.
- All viewing panels are stainless steel framed Acrylic.
- Front viewing panel is top hinged and self supporting for easy access to the work area for cleaning and equipment ingress.
- All filter diffuser/guard panels are removable for easy cleaning.
- Straight sides and back maximize work area to accommodate the many types and shapes of pharmacy equipment and compounders.
- All corners in work area, antechamber and work surface are easily reached and cleaned.
Antechamber
- Sealed two-door airlock (16"W x 13 1/2"D x 14 1/4"H) maintains complete environmental separation between work area and ambient conditions.
- HEPA filtered purge of trapped airlock air eliminates cross contamination between the work area and the room during both material ingress and egress.
Waste Removal
- Trash tube is a 4" diameter tube, flanged for waste bag/container connection and provided with an easily removable seal/stopper to maintain work area containment. (Stainless Steel trash can is available.)
- Sharps tube is a straight 2 1/2" diameter tube to facilitate quick drop of the largest syringes and is provided with an easily removable seal/stopper to maintain work area containment.
- Fully adjustable sharps container shelf accommodates most commercially available wide mouth or hinged top sharps collectors. (Examples: BD Part # 305610 and 305491)
Safety
- Lock-out handle requiring key for access to work area.
- Digital pressure readout with low pressure alarm for work area.
- Inward/Outward face velocity is 95-100 linear feet per minute at gloveport opening to protect operator/product during massive breach of containment.
- Glove changes can be made without breaking containment.
- Locking casters are standard, seismic anchors are available.
Ergonomics
- Hydraulic assist height adjusting stand offers a full 10" range variance allowing operators to sit or stand comfortably for extended time periods.
- Stainless steel sliding tray inside the airlock pulls into the work area eliminating reaching strains.
- Two part sleeve/glove system allows the use of most types and sizes of commercially available gloves for better dexterity and tactility.
- Extra large oval gloveports are placed with bottoms together to provide an anthropometrically correct configuration that accommodates a wide range of body types and increases range of movement.
- Gloveports have a 3" arm rest to reduce operator fatigue.
- Large viewing panel is at an angle to reduce glare and operator strain.
- Integrated foot rest.
Configuration Options
- Overall work area pressure can be either positive or negative and can be altered in place by a certifier.
- Antechamber can be left or right handed or can be installed on both sides.
- Waste and sharps removal tubes can be left or right handed or can be installed on both sides.
- Three available exhaust options that can be factory set or changed by a certifier in-place.
Filtration
- Full framed standard size minipleat HEPA/Filters filter 100% of inlet and exhaust air from both the work area and the antechamber to provide a fully contained environment.
- HEPA filters are full coverage and front loading for easy replacement by a certifier.
- A Pre-filter extends the life of the work area supply HEPA filter and can be replaced without tools.
Electrical
- High capacity motor/blower systems with speed control to extend HEPA filter's life.
- Ten foot hospital grade power cord with molded grounded plug.
- Sealed outlet in work area.
- Separate lighted power ON/OFF indicator switches for blower and lighting.
- High efficiency, standard sized fluorescent lights are externally mounted to minimize heat build-up.
- Voltage = 115 Volt, 60 Hz (220/50-60 Hz also available).
- Amperage Rating = 15
- Running Amperage = 6 (LFGI-3 and 4USP) =7(LFGI-6USP)
Exhaust
- Clean air plenum is external to the work area.
- Constructed to allow for optional outside venting of exhaust air.
- Constructed to allow for 3 exhaust options that can be factory set, or changed by a certifier in place.
- Venting is not required for operation but is optional.
| No Exhaust - 100% recycling of air | Total Exhaust - no recycling of air | |
| Exhaust Connection | N/A | 8" OD Flexible Duct |
| Exhaust CFM | 20 | 300 (LFGI-3 & 4USP) • 400 (LFGI-6USP) |
| Exhaust Static | N/A | Up to 2 inch w.c. (depending upon config) |
Weight
Net Weight :
|
Shipping Weight :
|
Section III: Optional Equipment and Accessories
(consult Your SalesPerson For Pricing)
- LFGI-A80 Ergonomic task chair with anti-bacterial and anti-microbial vinyl covering, various colors.
- LFGI-A85S Stainless steel cleaning handle with pivot head. Accepts lint free mop heads (LFGI-A85M).
- LFGI-A8E Electrical Height Adjustment (requires additional plug-in).
- LFGI-A83 Stainless steel garbage can (hangs on side to contain waste bags).
- LFGI training DVD and validation documents are also available.
Section IV: Unit Diagrams
LFGI-3USP
diagram

LFGI-4USP
diagram

LFGI-6USP
diagram

LFGI-6DUAL AIRLOCKdiagram |
![]() |
||
![]() |
|||
![]() |
|||
All units are made in the USA and available for purchase on the GSA.